1st BASE Magnetic Rack is designed for fast and efficient magnetic separation of magnetic beads in tube-based workflows, commonly used in nucleic acid extraction applications.

The magnetic rack is supplied with two-component system:

Magnet Rack (Base)

    • Permanent magnet with magnetic field intensity of more than 5500 gauss.


Tube Rack (Top)

    • Removable tube rack.

These components allow the tube rack to be easily detached from the magnetic base which provide conveniency to user. This feature enables direct placement of tubes on a shaker or vortex mixer during extraction procedures, without the need to remove the tubes from the rack.

//base-asia.com/wp-content/uploads/2026/03/KITMag--scaled.jpg

FEATURES

  • Strong permanent magnet
  • Efficient sample recovery
  • Enhanced sample visibility
  • Quick tube removal
  • Optimized tube angle for easy aspiration

APPLICATION DATA

Performance Review & Comparison Data

Magnetic Rack Performance

Visibility

    • Optimized design ensures high clarity, providing excellent visibility of magnetic bead binding during processing.
//base-asia.com/wp-content/uploads/2026/03/magnetic-rack.png

Figure 1. The rack designed for multi-angle clarity to ensure consistent observation of magnetic binding during procedures.

Magnetic Binding Time

    • 5-60 seconds (varies by magnetic bead solution), enabling rapid and efficient processing.


Bead Compatibility

    • Compatible with magnetic beads from multiple leading brands

Storage and Stability

    • The magnet, composed of permanent magnets, must be stored in a dry and cool place and should never be used at a temperature above 50oC.


Cleaning and Maintenance

    • Recommendation:
      • 70% Isopropyl Alcohol
      • 1% Sodium Hypochlorite Solution (bleach)
      • 0.1N HCI solution
    • Spray and wipe only - do not soak or immense into the cleaning solution
    • Avoid prolonged exposure to water or aqueous solution
    • DO NOT autoclave the rack

 

Product Code KIT-MAG16A
Product 1st BASE Magnetic Rack (1.5ml/2.0ml), for 16 tubes
Application Magnetic bead-based separation
Rack Dimension 216.9 mm (L) x 46.1 mm (W) x 47.1 mm (H)
Tube Capacity 16 tubes, Suitable for 1.5 mL/2.0 mL microcentrifuge tubes
Working Volume 0.1 mL – 2.0 mL
Operating Temperature  Magnetic rack should not be used above 50 ˚C